Fotona Replacement Part
Fotona LightWalker® Handpiece Replacement Mirror
SKU 88656 • Replacement Handpiece Bending Mirror
Maintain maximum optical efficiency and precise energy delivery with the SKU 88656 Replacement Handpiece Bending Mirror. Designed for Fotona Er:YAG dental handpieces (such as the H02 and H14 series), this high-precision internal mirror accurately reflects the 2940 nm laser beam through the handpiece head to ensure consistent clinical performance.
Key Features
◉
Optimal beam transmission: Restores full energy output and prevents beam attenuation caused by worn or damaged internal optics.
◈
Durable optical coating: Precision-engineered to withstand standard Er:YAG pulse characteristics.
✓
Easy maintenance: Can be inspected and carefully cleaned using specialized mirror tools and pure acetone or laboratory-grade alcohol.
Compatibility
Handpieces: Fotona Er:YAG dental handpieces, such as the H02 and H14 series
Wavelength: 2940 nm
Wavelength: 2940 nm
Why it matters
Water droplets or debris on the internal mirror surface can absorb laser energy and permanently damage the coating. Proper reprocessing protects your optics and keeps clinical performance consistent.
Sterilization & Reprocessing
1
Pre-cleaning handling: Wipe coarse debris off the handpiece body immediately after treatment. Hold the handpiece vertically when removing it from the articulated arm coupler to prevent water droplets from falling onto the internal optics.
2
Mandatory sealing before autoclaving: Dental Er:YAG handpiece bodies must be fully sealed before entering the autoclave. Insert the rear protective plug (locking it with a 90° turn) and ensure the front is sealed with a fiber tip, protective window, or metal cap to isolate the mirror chamber.
3
Autoclave parameters: Steam sterilize the sealed handpiece body at 132°C–134°C (270°F–273°F) for 3 to 8 minutes (or 15 minutes for gravity displacement cycles) with a minimum 30-minute drying period.
4
Post-sterilization precaution: Confirm the interior is completely dry before reattaching the handpiece body to the coupler. To safely clear any remaining microscopic moisture and protect the mirror, fire a few low-energy trial pulses before starting clinical treatment.
Never use flash, dry-heat, or chemical plasma sterilization.